FujiFilm CR-IR341 Service Manual page 353

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
03B4 Recovered IP Load Slow Conveyance Error
Cause
After the IP that experienced before-reading error is conveyed in fast speed mode to
the elevation unit, SB turns ON; the IP is then conveyed in slow speed mode toward
the cassette set unit, but SB2 remains ON. (When 03B4 error occurs, remedy the
before-reading error as well.)
Supplementary explanation
The IP is conveyed from the up-down IP removal unit to the side-positioning conveyor,
but before-reading error (IDT line error, etc.) occurs.
The IP that caused the before-reading error is conveyed (returned) from the side-
positioning conveyor to the up-down IP removal unit.
The IP conveyed is conveyed from the up-down IP removal unit toward the cassette by
the driving of MB2, and passes over SB2. (SB2: OFF to ON)
The IP is conveyed into the cassette and completely passes over SB2. (SB2: ON to
OFF)
This error occurs when the IP does not completely pass over SB2 (SB2 status does
not transition from ON to OFF) during the above operation.
Sensor related
Error-related I/O locations
Cassette inlet IP sensor
Error-related system block diagram
AC IN
DC OUT2
JPS-1
DC OUT3
009-051-04
05.31.99 FM2460
3
4
5
6
ON: Close, OFF: Open
SB2
SNS08A
CNB1
CN
CN13
CN4
DRV03D
CN3
CNB3
CN3
DRV08A
CN1
CR-IR341 Service Manual
MB2
IP transport motor
MC1
IP transport motor
FR1H0017.EPS
CNB2
CNMB2
MB2
FR1H0018.EPS
MT - 84
MT - 84

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr-ir341pFcr 5000Fcr 5000plus