FujiFilm CR-IR341 Service Manual page 433

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
2.5
IP Jam Handling Procedures
CAUTIONS
• Before removing the covers, be sure to turn OFF the high-voltage switch. If the machine is
powered ON with any of the covers removed, the photomultiplier will be damaged.
• Before returning a jammed IP to the machine, thoroughly make sure that the IP is not
scratched, scaled off, or seriously warped due to rubbing.
• Do not leave a jammed IP bent in the conveyance path for prolonged periods of time because
machine malfunction or IP breakage may occur.
• IP conveyance checkout must be performed after the machine is restored to normal with a
jammed IP removed.
Removing a Jammed IP
When an IP jam error occurs, you should basically complete the following procedure.
(1) Turn OFF the machine.
(2) Turn the machine back ON.
The initialization sequence is performed.
(3) Check whether the jammed IP is discharged in the initialization sequence.
When the IP is not discharged
Perform steps (4) and beyond.
When the IP is discharged
Check for improper IP conveyance.
"2.6 Checking for Improper IP Conveyance"
(4) Note the error code and remove the IP in a manner appropriate for the error encoun-
tered.
The tables on the subsequent pages indicate the relationship between the displayed
error codes and IP removal procedures. Note the tables to determine the IP removal
procedures to be performed. Remove the IP in accordance with the remedies (1
through 10) set forth in section 5.3.2, Detailed IP Removal Procedures.
(5) Restore the machine to normal.
(6) Check for improper IP conveyance.
"2.6 Checking for Improper IP Conveyance"
009-051-04
05.31.99 FM2460
3
4
5
6
CR-IR341 Service Manual
MT - 164
MT - 164

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr-ir341pFcr 5000Fcr 5000plus