FujiFilm CR-IR341 Service Manual page 734

Fuji computed radiography
Table of Contents

Advertisement

C
11
12
ONTENTS
12.15 Checking and Replacing Shock Absorber of Erasure
Conveyor
CAUTION
When removing and installing the units, use care not to damage the cables.
Checking and Replacing Shock Absorber on Before-Side-Positioning Conveyor Side
(1) Remove the erasure conveyor.
"12.2 Removing and Reinstalling Erasure Conveyor"
(2) Remove the shock absorber-attached guide plate.
(3) Check if the shock absorber is not peeled or frayed.
If it is peeled or frayed, replace the whole guide plate to which the shock absorber is
attached.
(4) Reinstall the guide plate in reverse order of removal.
009-051-04
05.31.99 FM2460
13
14
15
16
CR-IR341 Service Manual
(2), (3)
Shock absorber-attached
guide plate
Sems M3x6, x4
Nut, x4
Erasure conveyor
FR1H1623.EPS
MC - 243
MC - 243