FujiFilm CR-IR341 Service Manual page 409

Fuji computed radiography
Table of Contents

Advertisement

1
2
C
ONTENTS
13A3 Cassette Shelf Error
Cause
During cassette ejection for initialization or at the end of initialization, cassette hold
release is effected (SOLA1-4 are turned ON), but either SA3, 7, 11 or 15 fails to turn
OFF, so that the cassette hold release cannot be achieved.
Supplementary explanation
If there is a cassette during initialization:
During cassette ejection, SOLA1-4 are turned ON to effect cassette hold release.
As SOLA1-4 turn ON, the hold pin is lowered, so that SA3, 7, 11, and 15 are turned
OFF, thereby detecting the hold pin release.
This error occurs when either of the relevant shelves (SA3, 7, 11, 15) fails to turn OFF
so that the cassette hold release cannot be achieved during the above operation.
If there is no cassette during initialization:
During initialization, SOLA1-4 are turned ON to effect cassette hold release.
As SOLA1-4 turn ON, the hold pin is lowered, so that SA3, 7, 11, and 15 are turned
OFF, thereby detecting the hold pin release.
This error occurs when either of SA3, 7, 11, and 15 fails to turn OFF so that the
cassette hold release cannot be achieved during the above operation.
Reference: If all of SA3, 7, 11 and 15 do not turn OFF during the above operation, level 0
Sensor related
Error-related I/O locations
SOLF1
Error-related system block diagram
AC IN
DC OUT2
JPS-1
DC OUT3
009-051-04
05.31.99 FM2460
3
4
5
6
error "degeneration for all shelves" occurs.
ON: Close, OFF: Open
SA3, 7, 11, 15
Cassette hold sensor
SOLA1—4 ON :SA3, 7, 11, 15 OPEN
SOLA1
SOLA1—4 OFF :SA3, 7, 11, 15 CLOSE
SOLA2
Cassette hold solenoid
SOLA3
SOLA4
SF3
SNS08A
CNA2
CN
CN6
CN13
CN4
DRV03D
CN3
CNA3
CN3
SOLA2
CN4
DRV08A
SOLA3
CN1
SOLA4
CR-IR341 Service Manual
Hold pin
SOLA1—4
Cassette hold solenoid
SOLA1
FR1H0111.EPS
MT - 140
FR1H0110.EPS
MT - 140