FujiFilm CR-IR341 Service Manual page 370

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
03C0
Preliminary steps:
1. Power OFF the machine and remove the covers.
2. Turn OFF the high-voltage switch (S1) of the SCN08A board.
3. Check for any foreign matter (lead character marking chips, etc.) inside the machine, and, if any, remove it all.
Power OFF
Is the drive system of the side-
positioning conveyor normal?
Move the side-positioning conveyor
by hand to check that the latch
assembly pushes up the stopper.
Y
Power OFF
Is the IP normal?
Make sure that the IP is not curled,
bent, or warped. If it is curled, it
moves beyond the stopper and is
conveyed into the subscanning unit.
Check all the IPs as needed.
Y
Power OFF
Does the initialization self-diagnostics
end normally?
Y
Replace the SCN08A board.
009-051-04
05.31.99 FM2460
3
4
5
6
N
Adjust/replace the following parts.
N
Replace the IP(s).
N
CR-IR341 Service Manual
Belt/belt tension
Roller
Spring
Troubleshoot it based on a new error code displayed.
MT - 101
FR1H0044.EPS
MT - 101