5.14 Checking and Replacing Elevation Changeover Guide

Check and Replacement
(1) Check if the shock absorber of the elevation changeover guide is not peeled or frayed.
If it is peeled or frayed, proceed to step (2) to replace the whole elevation changeover
guide to which the shock absorber is attached.
(2) Remove the spring of the elevation changeover guide.
(3) Remove the single screw that retains the elevation changeover guide to the arm.
Using a Phillips stubby screwdriver or the like, remove the screw.
(4) Remove the arm.
(5) Remove both the right and left E-rings that retain the elevation changeover guide.
First remove the E-ring located on the left side of the machine.
(6) Remove both the right and left slide bearings.
Right side view of machine
(6)
Slide bearing
(5)
E-ring
009-051-04
05.31.99 FM2460
3
4
5
6
(1)
Shock absorber-attached
up-down changeover guide
CR-IR341 Service Manual
7
8
9
10
Vertical conveyor
Left side view of machine
MC - 82
(6)
Slide bearing
(5)
E-ring
(3)
Sems M3x10 for retaining
up-down changeover guide
(4)
Arm
(2)
Spring
FR1H1424.EPS