5.11 Replacing Up-Down Drive Motor (MB3)
(1) Rotate the MB3 timing belt to raise the up-down IP removal assembly to its topmost
position.
(2) Secure the IP removal unit in place with a cable tie so that it will not fall down.
CAUTION
Be sure to secure the IP removal unit in place. If not, theIP removal unit may fall down,
because the worm gear becomes free when MB3 is removed.
(2)
Cable tie
(3) Remove the cable clamp.
(4) Disconnect the connector (CNMB3).
(5) From the rear of the machine, remove the two MB3 retaining screws, and take out
MB3.
(6) Remove the timing belt wheel from the MB3 taken out, and attach it to a new MB3.
(7) From the rear of the machine, secure MB3 in place with two screws.
At this time, put the timing belt onto the timing belt wheel.
(8) Put the cable clamp back where it was.
(9) Connect the connector (CNMB3).
(10) Cut the cable tie that retains the IP removal unit in place.
009-051-04
05.31.99 FM2460
3
4
5
6
E-ring
(6)
Timing belt
wheel
Pan-head M4x10, x2
CR-IR341 Service Manual
7
8
9
10
(1)
Timing belt
IP removal unit
FR1H1072.EPS
Timing belt
(3), (4)
CNMB3
(5)
MB3
FR1H1073.EPS
MC - 76
MC - 76