Code
Name
2808
Output warning 1
2809
Output warning 2
No UID at time of DST output reservation
280B
designation
280C
LIF backup memory clear
2814
DSP10 checksum error
2815
DSP10 self-diagnostics error
2816
DSP10 image processing error
281B
DSP11 checksum error
281C
DSP11 self-diagnostics error
281D
DSP11 image processing error
2822
DSP20 checksum error
2823
DSP20 self-diagnostics error
2824
DSP20 image processing error
2829
DSP21 checksum error
009-051-04
05.31.99 FM2460
5
6
Significance/Occurrence Condition
For B4 film, there is a request for 4-image output.
At present, a 4-image output request is converted to a single
image, and nothing is done during output.
There is no UID that corresponds to the number of frames,
but it is ignored.
The backup memory is cleared, and nothing is done.
DSP checksum error (minor).
DSP diagnostics error (minor).
DSP image parameter error (minor).
DSP checksum error (minor).
DSP diagnostics error (minor).
DSP image parameter error (minor).
DSP checksum error (minor).
DSP diagnostics error (minor).
DSP image parameter error (minor).
DSP checksum error (minor).
CR-IR341 Service Manual
Detail: Denotes the detail information format.
Probable Cause/Remedy
Nothing is done. At present, after all the other films
under request are outputted, it is converted to a single
image for output.
Nothing is done.
It should be ignored.
Nothing is done.
Check the IMG08A board.
Check the IMG08A board.
Check the image parameter.
Check the IMG08A board.
Check the IMG08A board.
Check the image parameter.
Check the IMG08B board.
Check the IMG08B board.
Check the image parameter.
Check the IMG08B board.
MT - 37
Detail
Refer to
I-6
I-6
I-10
I-7
I-7
I-7
I-7
I-7
I-7
I-7
I-7
I-7
I-7
TR1H1251.EPS