03B6 Feed/Load Conveyance Remaining IP Ejection Error
Cause
During initialization, an IP is detected within the machine as a result of IP search, and
it is loaded to the cassette, but SB2 remains OFF, or it turns ON but does not turn
OFF. (When 03B6 occurs, the cause of the IP jam should be checked and remedied.)
Supplementary explanation
During initialization, each motor is activated, and each sensor is used to search for any
remaining IP left in the machine.
If any IP is detected in the machine, the IP ejection process is executed.
The IP is conveyed toward the cassette by the driving of MF1, MC1, and MB2.
The IP is conveyed from the up-down IP removal unit toward the cassette, and passes
over SB2. (SB2: OFF to ON)
This error occurs when the IP fails to pass over SB2 (SB2 status does not transition
from OFF to ON), or when it does not completely pass over SB2 (SB2 status does not
transition from ON to OFF), during the above operation.
Sensor related
Error-related I/O locations
SB2
Cassette inlet IP sensor
Error-related system block diagram
AC IN
DC OUT2
JPS-1
DC OUT3
009-051-04
05.31.99 FM2460
3
4
5
6
ON: Close, OFF: Open
SNS08A
CN
CN13
CN4
DRV03D
CN3
CN3
CNC2
DRV08A
CN1
CR-IR341 Service Manual
MB2
IP transport motor
MF1
IP transport motor
MC1
IP transport motor
SF1
MZ1
ME2
CNB1
CNB2
CNMB2
CNB3
MB2
CNMC1
MC1
FR1H0024.EPS
MT - 88
MD3
FR1H0023.EPS
MT - 88

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr-ir341pFcr 5000Fcr 5000plus

Table of Contents