03D3 Before-Erasure Conveyance Error
Cause
After the IP is conveyed into the inverting stocker, it is conveyed toward the erasure
conveyor, but SF2 does not turn ON.
Supplementary explanation
The IP conveyed from the after-reading conveyor is conveyed into the inverting stocker
of the erasure conveyor.
After the IP passes over SF2 (SF2: ON to OFF), SOLF1 is driven and the path
changeover guide is activated.
As the path changeover guide is activated, the IP conveyance direction is switched to
the erasure lamp unit.
After the conveyance direction is switched, the IP is conveyed toward the erasure lamp
by the driving of MF1, and the IP passes over SF2 again. (SF2: OFF to ON)
This error occurs when the IP fails to pass over SF2 (SF2 status does not transition
from OFF to ON) after the activation of the path changeover guide during the above
operation.
Sensor related
Error-related I/O locations
IP transport motor
SF4
Overrun IP sensor
Changeover IP sensor
Error-related system block diagram
AC IN
DC OUT2
JPS-1
DC OUT3
009-051-04
05.31.99 FM2460
3
4
5
6
ON: Close, OFF: Open
MF1
SF2
SNS08A
CN
CN8
CN13
CN4
DRV03D
CN3
CN3
CN10
DRV08A
CN1
CR-IR341 Service Manual
SOLA3
SOLA4
FR1H0068.EPS
CNF1
CF4
CNF2
MF1
CNMF1
FR1H0069.EPS