(4) Connect the connectors (CN1, 2) to the PMT08A board.
PMT08A board
(5) Power ON the machine.
(6) Touch the U-Utility button.
The screen switches to the U-Utility mode.
(7) Touch the upper left corner and upper right corner of the operation panel in sequence.
M-Utility is activated.
(8) Select "5. SCANNER UTILITY" and "5. HV DATA" in sequence.
(9) Type in "250" as a high-voltage setup value.
The high voltage is set to 250 V.
(10) With the terminals (HV135, PGND) touched by the probes of the tester, turn ON the
high-voltage switch (S1) of the SCN08A board.
CAUTION
When making voltage measurements on the PMT08A board, be sure to touch the
terminals of the board by the tester probes with the high-voltage switch in its OFF
position, and then turn ON the high-voltage switch. It is dangerous if the high-voltage
switch is turned ON, because high voltage is applied to the PMT08A board.
(11) Measure the voltage between the terminals of the PMT08A board.
HV135-PGND: -17.8 ± 2V
PMT08A board
009-051-04
05.31.99 FM2460
3
4
5
6
CN1
CR-IR341 Service Manual
7
8
9
10
CN2
FR1H1537.EPS
(11)
CN2
HV135(TP5)
PGND
FR1H1538.EPS
MC - 177
MC - 177