Replacement of tensioner
(1) Remove the following covers.
Lower front cover, lower right-hand side cover
"3. Removing and Reinstalling Covers"
(2) Remove the transparent cover located on the right-hand side.
Using a blower, remove dust deposited on the transparent cover.
(3) Remove the tensioner spring located on the right-hand side.
(4) Replace the SUS belt.
(5) Replace the tensioner.
(6) For reinstallation, reverse the removal steps.
NOTE
After installing the right-hand side tensioner and SUS belt, rotate the drive shaft to
check that the SUS belt or rubber belt does not come off the pulley.
(7) Power ON the machine.
(8) Check MZ1.
Make sure that the motor rotates normally.
"10.8 Checking and Replacing Subscanning Motor (MZ1)"
(9) Check image performance.
"3.2 Checking Image" in Preventive Maintenance Volume
009-051-04