C
1
2
ONTENTS
(2) Detailed Sequence of Reverse Conveyance Load
It is the detailed sequence of reverse conveyance load for "Error Handling: Recovered IP
Load."
Fig. 6.3.2 Reverse Conveyance Load
Preprocessing: Fig. 6.3.1, Fig. 6.7.1, Fig. 6.8.1, and Fig. 6.9.1
START
N
Y
Recovered IP load slow conveyance stop
DRIVE (MD3/PD81)
DRIVE (MC2/PC12)
DRIVE (MB2/PB23)
N
Y
Recovered IP load slow conveyance stop
DRIVE (MD3/OFF) (soft)
DRIVE (MC2OFF) (soft)
DRIVE (MB2/OFF) (hard)
END
Table 6.3.2 Operation Parameters
Symbol
PB23
Conveyance (after-erasure conveyance)
PC12
Conveyance (after-erasure conveyance)
PD23
Conveyance grip (HP return)
PD81
Recovery conveyance (fast load conveyance)
TB22
Load timeout
TB81
Recovery load timeout
009-051-04
05.31.99 FM2460
3
4
5
6
Recovered IP load fast conveyance
DRIVE (MD3/PD81)
DRIVE (MC2/PC12)
DRIVE (MB2/PB23)
Is SB2 closed within TB81?
Is SB2 opened within TB22?
Load suction
Fig.2.10.3
Load suction conveyance
Fig.2.10.4
Conveyance grip (HP return)
DRIVE (MD2/PD23)
Description
CR-IR341 Service Manual
7
8
9
10
Recovered IP load fast conveyance stop
DRIVE (MD3/OFF) (soft)
DRIVE (MC2OFF) (soft)
DRIVE (MB2/OFF) (soft)
MD3, MC2, MB2, excitation OFF
Level 0 error
Conveyance error
03B5
(recovered IP load fast)
Recovered IP load slow conveyance stop
DRIVE (MD3/OFF) (soft)
DRIVE (MC2OFF) (soft)
DRIVE (MB2/OFF) (soft)
MD3, MC2, MB2, excitation OFF
Level 0 error
03B4
Conveyance error
(recovered IP load slow)
If reverse conveyance
load is performed for
any out-of-spec IP,
conveyance grip
(HP return) is not
executed.
Value
- infinite p
+ infinite p
400 p
+ infinite p
2.9 sec
5.1 sec
TR1H0100.EPS
MD - 157
FR1H0444.EPS
MD - 157