FujiFilm CR-IR341 Service Manual page 108

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
7.11 Power Supply
The power supply unit provides power supplies of +5VDC, +12VDC, +15VDC, and +24VDC
to the respective boards.
The power supply unit contains an inverter power supply circuit for control of lamp turn-ON,
thereby allowing the erasure lamp to turn ON.
For protection of the power supply circuitry, the output of the power supply unit is discon-
nected upon detection of overvoltage or overcurrent.
For recovery, manually turn OFF the breaker and back ON again.
The details of the outputs are described below.
Output
terminal
CN4
TB2
DCOUT1
DCOUT2
DCOUT3
DCOUT4
DCOUT5
DCOUT6
DCOUT7
009-051-04
05.31.99 FM2460
3
4
5
6
Output
Voltage
voltage
accuracy
+5V
– 5%
+5V
– 5%
+12V
– 5%
+5V
– 5%
+15V
– 5%
+24V
– 5%
+24V
– 5%
+24V
– 5%
+24V
– 5%
+24V
– 5%
+24V
– 5%
+5V
– 5%
+24V
– 5%
+15V
– 5%
-15V
– 5%
+5V
– 5%
+15V
– 5%
CR-IR341 Service Manual
7
8
9
10
Variable
Output current
voltage
Minimum
Steady state
range
0.1A
Fixed
– 10%
2.0A
– 10%
1.0A
0.5A
Fixed
– 10%
0A
– 10%
0.52A
– 10%
0A
– 10%
0A
– 10%
0A
0A
Fixed
– 10%
0A
– 10%
1.0A
– 10%
0A
– 10%
0.1A
– 10%
0.1A
0A
Fixed
– 10%
0A
MD - 45
Maximum
destination
0.1A
0.1A
LED08A
37A
37A
MTH08A
9.5A
9.5A
MTH08A
3.0A
3.0A
SNS08A
0.6A
2.4A
DRV08A
2.77A
3.1A
2.77A
3.1A
2.77A
3.1A
2.77A
3.1A
8.0A
12A
LAMP
0.26A
0.26A
LAMP
FAN
3.2A
3.2A
SCN08A
0.3A
0.6A
2.15A
3.0A
0.95A
3.0A
0.1A
0.1A
FFM
0.4A
2.0A
MD - 45
Output
TR1H0004.EPS

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr-ir341pFcr 5000Fcr 5000plus

Table of Contents