FujiFilm CR-IR341 Service Manual page 722

Fuji computed radiography
Table of Contents

Advertisement

C
11
12
ONTENTS
12.8 Replacing Erasure Lamp
(1) Remove the erasure lamp assembly.
"12.2 Removing and Reinstalling Erasure Conveyor"
(2) Remove the bracket.
(3) Remove the five erasure lamps.
(4) Wipe the white reflection face within the erasure lamp assembly and the white reflec-
tion face behind the bracket with a dry cloth.
(3)
Erasure lamp x5
(4)
Reflection
face
(5) Install five new erasure lamps.
(6) Attach the bracket.
(7) Reinstall the erasure lamp assembly.
"12.2 Removing and Reinstalling Erasure Conveyor"
009-051-04
009-051-04
05.31.99 FM2460 (1)
05.31.99 FM2460
13
14
15
16
(4)
Reflection
face
CR-IR341 Service Manual
(2)
Bracket
Sems M3x6 x2
FR1H1266.EPS
MC - 231
MC - 231

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr-ir341pFcr 5000Fcr 5000plus

Table of Contents