(3) Sequence of Erasure Conveyance
In this sequence, the IP is erased and fed to the IP removal conveyor.
After this sequence, the erasure lamp turn-off process is performed. For more detail, see
Section 2.9 (1).
Fig. 2.9.3 Erasure Conveyance
2w-1-2 phase
Elevation unit
movement (load)
2w-1-2 phase
Table 2.9.3 Operation Parameters
Symbol
PF13
Conveyance (erasure 1)
PF14
Conveyance (erasure 2)
TF15
Erasure conveyance IPtrailing-edge
detection timeout
009-051-04
05.31.99 FM2460
3
4
5
6
Preprocessing: Fig. 2.9.1
START
Is SF2 opened?
Y
N
Erasure conveyance
DRIVE (MF1/PF13)
Is SF2 opened within TF15?
N
Y
Erasure conveyance stop
DRIVE (MF1/OFF)
(soft)
Specified erasure conveyance
DRIVE (MF1/PF14)
MF1 excitation
stability point reset (2-phase)
END
Value
Description
+ infinite p
15,992 p
Variable
depending
on exposure
TR1H0041.EPS
CR-IR341 Service Manual
7
8
9
10
Y
N
Erasure conveyance stop
DRIVE (MF1/OFF) (soft)
MF1 excitation OFF
03D5
IP path
FRONT
Erasure conveyor
MD - 96
Is SF3 closed, as the conveyance
of PF14 is performed with erasure
conveyance?
Level 0 error
Conveyance error
(erasure)
FR1H0372.EPS
IP inverting stocker
IP path
Erasure lamp
Path changeover
mechanism
MD - 96
FR1H0373.EPS