Preliminary steps:
1. Power OFF the machine and remove the covers.
2. Turn OFF the high-voltage switch (S1) of the SCN08A board.
3. Check for any foreign matter (lead character marking chips, etc.) inside the machine, and if any, remove it all.
Power OFF
Is the IP normal?
Make sure that its sucked surface
is not soiled, curled, or bent.
Y
Power OFF
Are the suction cups of the IP
removal unit normal?
Check if the suction cups are
installed properly.
Check the suction cups for any
soil or deformation.
Y
Power OFF
Is the drive mechanism of the IP
removal unit normal?
Move the suction cup arm by
hand to check that it moves
smoothly.
Y
Power ON
Does PB1 rotate along during
initialization self-diagnostics?
Y
Power ON
Is SB5 normal?
With "6. MECH" of M-Utility,
execute sensor monitoring to
check the SB5 status.
Y
Power OFF
Is SB5 installed normally?
Power OFF
Is the connector connection
from SB5 to SNS08A board
normal?
Restart the machine
Replace the following parts
and check its status.
in order named.
SB5
SNS08A board
009-051-04
05.31.99 FM2460
3
4
5
6
N
Clean/replace the IP.
N
Clean/adjust/replace the following parts.
Suction cup rubber
Suction cup hose
N
Adjust/replace the following parts.
Suction cup arm/shaft/bearing
Power OFF
N
Are the connector
connections from PB1 to
DRV08A to JPS-1 board normal?
Y
Replace the following parts
in order named.
PB1
DRV08A board
N
N
Install the sensor properly.
Y
N
Check the error-related
system block diagram
and connect the connectors
Y
as appropriate.
CR-IR341 Service Manual
N
Check the error-related
system block diagram
and connect the connectors
as appropriate.
FR1H0136.EPS
MT - 163
MT - 163