FujiFilm CR-IR341 Service Manual page 637

Fuji computed radiography
Table of Contents

Advertisement

1
2
C
ONTENTS
(3) Pull out the scanner unit.
"7 Pulling Out and Pushing In Scanner Unit"
(4) Make sure that the connectors (CN1-4) of the LDD08A board are locked.
If not, securely connect the connectors to lock them in place.
CAUTION
When servicing the LD board, be sure to wear an anti-static wristband to ensure
proper grounding. If not, static electricity built on the human body may cause damage
to electric parts mounted on the board.
(4)
CN2 CN3
CN4
(5) Push in the scanner unit.
"7. Pulling Out and Pushing In Scanner Unit"
(6) Make sure that connector (CN1) of the SCN08A board is locked.
If not, securely connect the connector to lock it in place.
CAUTION
When connecting CN1, use care not to touch the C33 and C94 capacitors.
(7) Make sure that the no-fuse breakers (NFB1, 3) of the SCN08A board are in their
shorted condition (button depressed).
Push the no-fuse breaker button several times to place the breaker in its shorted
condition.
NOTE
With board version J or later, the no-fuse breaker (NFB) is replaced with the fuse.
Thus, make sure that the fuse has not been blown.
009-051-04
009-051-06
04.20.2001 FM3006
05.31.99 FM2460
3
4
5
6
Scanning optics unit
Protective
top cover
LDD08A board
CN1
Top view
CR-IR341 Service Manual
7
8
9
10
(4)
CN3
CN2
CN4
LD assembly cover
Sponge
Side view
MC - 146
Protective top cover
LDD08A board
CN1
Optics base
FR1H1276.EPS
MC - 146