11.6 5. UNIT: 2. UP-DOWN UNIT: Operation Tests on Elevation
Unit
1. HOME POSITION: To Locate HP of IP Removal Unit
Execute "1. HOME POSITION".
The following message appears.
UP/DOWN UNIT H.P. SETTING IS IN PROGRESS.
RESULT - OK (or RESULT - ERROR)
Meaning: IP removal unit home positioning is now being performed.
2. MOVE POSITION: To Input Unit Position
(1) Select "2. MOVE POSITION".
The following message appears.
INPUT THE UNIT POSITION.
1 - 4 :
Meaning: Enter the unit position.
(2) Type in the unit position.
The following message appears.
MOVED-UP/DOWN IS IN PROGRESS.
RESULT - OK (or RESULT - ERROR)
Meaning: The elevation unit is now being moved.
009-051-05
11.20.99 FM2638
13
14
15
16
CR-IR341 Service Manual
17
MU - 100
MU - 100