FujiFilm CR-IR341 Service Manual page 306

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
3
4
ONTENTS
Code
Name
2A11
Menu selection method inconsistent
2A20
Patient information data error
2A21
Undefined command received
2A22
Unexpected command received
2A23
Serial line test error
2A24
No response from external unit
Patient information search error
2A25
(not found)
2A26
Patient information search timeout
2A30
Patient information data error
2A31
Card read error
Card without relevant search key data
2A32
item
009-051-04
05.31.99 FM2460
5
6
Significance/Occurrence Condition
Although the routine menu display setup file is absent, the
"menu selection method" stored in the backup memory is
"routine menu selection."
Improper data is set in the patient information sent from the
online-connected unit.
A command that is not defined in the interface specification,
or an unimplemented command, is received from the
online-connected unit.
An unexpected command is received from the online-
connectedunit.
An error occurs in the serial line test with the online-
connected unit.
A response to the command requested in the communication
with the online-connected unit does not arrive within a certain
time.
A response indicating that patient information relevant to a
search key is not found is received from the type B online-
connected unit.
A response indicating that a timeout occurs in the search for
patient information based on a search key is received from
the type B online-connected unit.
Improper data is set in the patient information data read by
the magnetic card reader.
Improper data is set in the patient information data read by
the magnetic card reader.
The card read by the magnetic card reader does not contain
search key data specified by the user.
CR-IR341 Service Manual
Detail: Denotes the detail information format.
Probable Cause/Remedy
HDD, backup memory
External unit (online-connected unit), CPU90E board
External unit (online-connected unit), CPU90E board
External unit (online-connected unit), CPU90E board
External unit (online-connected unit), CPU90E board,
line cable
External unit (online-connected unit), CPU90E board,
line cable
External unit (online-connected unit)
External unit (online-connected unit)
Magnetic card, magnetic card reader, SNS08A board
Magnetic card, magnetic card reader, SNS08A board
Inconsistency between the magnetic card and search
key type selection (operation error)
MT - 43
Detail
Refer to
K-2
K-3
K-4
K-4
K-2
K-4
K-2
K-2
K-5
K-6
K-7
TR1H1257.EPS
MT - 43

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr-ir341pFcr 5000Fcr 5000plus

Table of Contents