FujiFilm CR-IR341 Service Manual page 194

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
(12) Detailed Sequence of Movement of Before-Reading Conveyance Remaining IP
Fig. 5.3.12 Movement of Before-Reading Conveyance Remaining IP
Preprocessing: Fig. 5.3.1
Initialization
side-positioning
conveyance
Table 5.3.12 Operation Parameters
Symbol
Initialization conveyance (DOWN direction)
PD71
Initialization conveyance (side-positioning conveyance reverse direction)
PD73
TD72
Side-positioning conveyance remaining IP ejection timeout 2
Wait for side-positioning switchback
TD74
Initialization side-positioning timeout
TD75
009-051-04
05.31.99 FM2460
3
4
5
6
START
Side-positioning conveyance reverse direction
DRIVE (MD3/PD73)
DRIVE (MZ1/ON) Fast reverse conveyance
Is SC1 closed within TD72?
N
Y
Side-positioning conveyance
reverse direction stop
DRIVE (MD3/OFF) (soft)
DRIVE (MZ1/OFF) (soft) reverse
DELAY TD74
Side-positioning conveyance direction
DRIVE (MD31/PD71)
Is SD3 closed within TD75?
N
Y
Side-positioning conveyance
direction stop
DRIVE (MD31/OFF)
(soft)
Before-reading standby IP found
END
Description
CR-IR341 Service Manual
7
8
9
10
Side-positioning conveyance
reverse direction stop
DRIVE (MD3/OFF) (soft)
DRIVE (MZ1/OFF) (soft) reverse
Level 0 error
Side-positioning conveyance
03B7
remaining IP ejection error
Side-positioning conveyance
reverse direction stop
DRIVE (MD31/OFF)
(soft)
Level 0 error
03B7
Side-positioning conveyance
remaining IP ejection error
Value
+ infinite p
- infinite p
5.9 sec
0.5 sec
5.8 sec
TR1H0069.EPS
MD - 131
FR1H0416.EPS
MD - 131

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr-ir341pFcr 5000Fcr 5000plus

Table of Contents