(4) Select "7. FILE UTILITY".
0. QUIT
1. ERROR LOG UTILITY
2. CONFIGURATION SETTING
3. TEST MODE
4. ELECTRICAL UTILITY
5. SCANNER UTILITY
6. MECHANICAL UTILITY
7. FILE UTILITY
8. BACKUP MEMORY
9. HV OFF
10. MENU SETTING
> 7
(5) Select "3. BACK UP" and "2. CONFIGURATION DATA" and "1.YES" in sequence.
> 7
0.QUIT
1.FORMAT FD
2.FORMAT IMAGE PARTITION
3.BACK UP
4.RESTORE
5.EDR DATA
6.PREVIOUS SYSTEM SOFTWARE
7.EXECUTION
FUT> 3
0.QUIT
1.SCANNER DATA
2.CONFIGURATION DATA
3.NETWORK DATA
4.IMAGE PROCESSING DATA
5.CSL MENU DATA
FUT:RES>2
NOTE
The menu parameters saved in the floppy disk also contain the language setting for the
machine. Thus, if the language setting differs during RESTORE and during BACK UP,
the file cannot be read during RESTORE.
009-051-05
11.20.99 FM2638
3
4
5
6
Of the setup files, the following files are saved to the floppy disk.
• SYSTEM (IRSET.CFG,IRSET.ORG)
• PRINT (FILFMT.CFG,FILFMT.ORG)
• IRSTATUS (IRSTATUS.CFG,IRSTATUS.ORG)
CR-IR341 Service Manual
7
8
9
10
A
B
D
E
7
8
4
5
1
2
0
.
DEL
BS
A
B
D
E
7
8
4
5
1
2
0
.
DEL
BS
MU - 64
C
F
9
6
3
SP
ENT
Caps
FR1B1200.EPS
C
F
9
6
3
SP
ENT
Caps
FR1B1214.EPS
MU - 64

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr-ir341pFcr 5000Fcr 5000plus

Table of Contents