03B7 Side-Positioning Conveyance Remaining IP Ejection Error
Cause
During initialization, a remaining IP is detected in the side-positioning conveyor or
subscanning unit, and it is conveyed toward the before-side-positioning conveyor by
the driving of MD3/MZ1, but SC1 does not turn ON. Or, after the IP trailing edge is
detected by SZ1, the IP is conveyed to the side-positioning conveyor, but SD3 does
not turn ON. (When 03B7 occurs, the cause of the JP jam should also be remedied.)
Supplementary explanation
During initialization, each motor is activated, and each sensor is used to search for any
remaining IP left in the machine.
A remaining IP is detected by SD3/SZ1 of the side-positioning conveyor.
In order to execute the IP ejection process, the IP is conveyed to the before-side-
positioning conveyor by the driving of MD3/MZ1.
When the IP conveyed to the before-side-positioning conveyor is detected by SC1, the
conveyance direction is switched to the side-positioning conveyor.
When SD3 of the side-positioning conveyor detects the IP, the IP is conveyed again
toward the before-side-positioning conveyor, so that the ejection process is executed.
This error occurs when SC1 fails to detect the IP, or when SD3 fails to detect the IP
after SC1 detects it, during the above operation.
Sensor related
Error-related I/O locations
Error-related system block diagram
AC IN
DC OUT2
JPS-1
DC OUT3
DC OUT7
009-051-04
05.31.99 FM2460
3
4
5
6
ON: Close, OFF: Open
SC1
Before-BCR IP sensor
SNS08A
CNF1
CN
CN8
CN13
CNC1
CN4
CN9
DRV03D
CN3
CNZ2 CND1
CNZ7
CN3
CN13
CNMD3
DRV08A
CN1
CN14
CNZ4
CR-IR341 Service Manual
SOLA4
MD3 IP transport motor
MD3
MZ1
FR1H0027.EPS
MT - 90
SD3
Side-positioning
IP sensor
FR1H0026.EPS