(9) Rotate the timing belt wheel by hand until the actuator blocks the SB1 light path.
A message indicating that the sensor is CLOSE appears on the operation panel.
If anything abnormal is found, such as no message indicative of the sensor OPEN or
CLOSE status displayed, then replace the sensor.
(10) Repeatedly select "0. QUIT".
M-Utility is exited.
Replacement
(1) Remove the cassette set unit.
"4.2 Removing the Reinstalling Cassette Set Unit"
(2) Disconnect the connector (SB1).
NOTE
If it is difficult to disconnect the connector, rotate the MB3 timing belt to move the IP
removal unit to such a position as to facilitate its disconnection.
When moving the IP removal unit, locate the IP removal arm in its home position to
prevent damage to it.
(3) Pull the IP removal arm toward yourself.
(4) Remove the suction cup HP sensor (SB1).
"3.5 Sensors Not Requiring Adjustment of Its Mounting Position" in Machine
Description
(5) Install a new sensor.
(6) Connect the connector (SB1).
(7) Check SB1.
" Checks"
009-051-04
05.31.99 FM2460
3
4
5
6
CR-IR341 Service Manual
7
8
9
10
IP removal unit
SB1
FR1H1312.EPS
MC - 61
MC - 61