11.4 Replacing IP Transport Motor (ME2)
The IP transport motor (ME2) must be replaced with the scanner unit pulled out.
(1) Remove the after-reading conveyor.
"11.1 Removing and Reinstalling After-Reading Conveyor"
(2) Disconnect the connector (CNME2).
(3) Remove ME2.
(4) Remove the timing belt wheel from ME2.
After-reading conveyor
(4)
Timing belt wheel
Pan M4x10 x3
CNME2
(5) Install a new ME2 by reversing the removal steps.
009-051-04