Adjustment
CAUTION
Before removing the covers, be sure to turn OFF the high-voltage switch. If the machine is
powered ON with any of the covers removed, the photomultiplier will be damaged.
(1) Remove the shutter assembly.
"4.3 Removing and Reinstalling Shutter Assembly"
(2) Remove the retaining bracket used for transit, which is located on the rear of the up-
down IP removal unit.
(3) Loosen the three mounting screws so as to allow the solenoid mounting location to be
shifted.
(4) Remove the spring.
(5) Put the protruding portion of the retaining bracket used for transit into the cassette hold
pin sliding hole, and then grip the retaining bracket used for transit with the cover-
closing roller.
(6) Raise the solenoid until the cassette hold pin contacts the protruding portion of the
retaining bracket used for transit, and then secure the solenoid in place.
The solenoid should be raised so that the retaining bracket used for transit is not lifted
up.
(7) Remove the retaining bracket used for transit.
(8) Push down the iron core of the solenoid by hand.
Make sure that the cassette hold pin is located below the guide plate of the cassette
set unit.
(9) Reinstall the spring.
(10) Check the solenoid adjusted.
" Checks"
009-051-04
05.31.99 FM2460
3
4
5
6
Cassette set unit
CR-IR341 Service Manual
7
8
9
10
(5), (6)
Hole
SOLA1
(6)
SOLA2
SOLA3
SOLA4
SOLA1, 2, 3, 4
MC - 46
Guide plate of
cassette set unit
Cover-closing
roller
Protruding portion of
retaining bracket used
for transit
Retaining bracket used for
transit
Hold pin
(4)
Spring
FR1H1108.EPS
MC - 46