10.9 Replacing Driving Shaft Grip Motor (MZ2)
Removal
(1) Pull out the scanner unit.
"7. Pulling Out and Pushing In Scanner Unit"
(2) Remove the after-reading conveyor.
"11.1 Removing and Reinstalling After-Reading Conveyor"
(3) Remove the FAST-ON terminals (MZ2(+), MZ2(-)).
(4) Remove the right-hand side transparent cover.
(5) Remove the SZ2-attached bracket as a whole.
(6) Remove the ring and cam from the MZ2 shaft.
(7) Remove MZ2.
(8) Reverse the removal steps to reinstall a new MZ2.
When installing the MZ2 bracket, screw it down while pressing it against the position-
ing bracket.
009-051-04
05.31.99 FM2460
3
4
5
6
(4)
Right-hand side transparent cover
BR M4x8, x2
CR-IR341 Service Manual
7
8
9
10
Subscanning unit
Positioning bracket
(6)
Ring
Hex socket head setscrew
M3x6
(5)
Bracket
(6)
BR M4x8, x4
Cam
SZ2
MC - 190
(7)
MZ2
Sems M3x6, x3
(3)
FAST-ON terminal (+)
(3)
FAST-ON terminal (-)
FR1H1020.EPS