Removing And Reinstalling Before-Side-Positioning Conveyor - FujiFilm CR-IR341 Service Manual

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
6.
Before-Side-Positioning Conveyor
This section describes procedures for removing and reinstalling the before-side-positioning
conveyor, and procedures for checking and replacing parts of the before-side-positioning
conveyor to be replaced.
Unless otherwise specified, it is assumed that covers that should be removed prior to checks
and replacement have been removed.
"3. Removing and Reinstalling Covers"
6.1
Removing and Reinstalling Before-Side-Positioning
Conveyor
Removal
(1) Shift the two tensioners until the timing belt can be removed.
Loosen the two screws (BR M4x8) that retain the tensioner to slide the tensioner as
appropriate, and then secure the tensioner in place.
"3.4 Tensioner Not Requiring Tension Adjustment" in Machine Description
(2) Remove the two timing belts.
(3) Disconnect the connectors (CNC1, CNC2, CNSOLC1).
(4) Remove the two screws that retain the before-side-positioning conveyor
(5) Pull out the before-side-positioning conveyor horizontally.
Slowly pull out the unit so that it is not caught by the connectors (CNC1, CNC2,
CNSOLC1).
(1), (2)
Timing belt
Erasure conveyor
009-051-04
05.31.99 FM2460
3
4
5
6
Tensioner
BR M4x8
CNC1
Before-side-
positioning conveyor
(4)
Before-side-
positioning conveyor
BR M4x8, x2
(3)
CNC1
(3)
CNC2
(3)
CNSOLC1
CR-IR341 Service Manual
7
8
9
10
(1), (2)
BR M4x8
positioning conveyor
(5)
MC - 86
Up-down IP removal
Timing belt
Before-side-
Tensioner
FR1H1109.EPS
MC - 86

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr-ir341pFcr 5000Fcr 5000plus

Table of Contents