C
11
12
ONTENTS
11.8 5. UNIT: 4. SIDE-POSITIONING UNIT: Operation Tests on
Side-Positioning Mechanism
1. HOME POSITION: To Locate HP of Side-Positioning Mechanism
Execute "1. HOME POSITION".
The following message appears.
MOVED TOWARD THE HOME POSITION IS IN PROGRESS.
RESULT - OK (or RESULT - ERROR)
Meaning: Side-positioning mechanism home positioning is now being performed.
2. ACTUATION: To Effect Side-Positioning Operation
(1) Select "2. ACTUATION".
The following message appears.
1.NORMAL 2.STEP :
(2) Enter the side-positioning operation mode (either "1" or "2").
When "1. NORMAL" is selected
When a series of side-positioning operations is initiated, the following message
appears.
SIDE-POSITIONING PROCESS IS IN PROGRESS.
ARE YOU SURE TO MOVE ON NEXT STEP?
Meaning: Side-positioning operation is now being performed.
If it ends in error, "ERROR (XXXX)" is displayed. "XXXX" denotes an error code.
When "2. STEP" is selected
When the side-positioning operation is initiated in steps, the following message
appears.
SIDE-POSITIONING PROCESS IS IN PROGRESS.
ARE YOU SURE TO MOVE ON NEXT STEP?
Meaning: Are you sure to proceed to the next step?
Hit the ENT key to proceed to the next step (there are four steps).
When the operation is completed, the following message appears.
RESULT - OK
If it ends in error, "ERROR (XXXX)" is displayed. "XXXX" denotes an error code.
009-051-05
11.20.99 FM2638
13