(7) Select "7. FILE UTILITY", "4. RESTORE" and "5. CSL MENU DATA" in sequence.
7. FILE UTILITY
8. BACKUP MEMORY
9. HV OFF
10. MENU SETTING
> 7
0.QUIT
1.FORMAT FD
2.FORMAT IMAGE PARTITION
3.BACK UP
4.RESTORE
5.EDR DATA
6.PREVIOUS SYSTEM SOFTWARE
7.EXECUTION
FUT> 4
0.QUIT
1.SCANNER DATA
2.CONFIGURATION DATA
3.NETWORK DATA
4.IMAGE PROCESSING DATA
5.CSL MENU DATA
FUT:RES>5
(8) Remove the "Menu default FD for USA" setup file floppy disk from the FD drive.
(9) Select "0. QUIT" repeatedly.
The screen returns to the U-Utility mode.
(10) Press the RESET button.
The machine is reset so that the setting of the file installed becomes effective.