Replacing Elevation Timing Belt - FujiFilm CR-IR341 Service Manual

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS

5.12 Replacing Elevation Timing Belt

(1) Rotate the MB3 timing belt to raise the IP removal unit to its topmost position.
(2) Secure the up-down IP removal assembly in place with a cable tie so that it will not fall
down.
CAUTION
Be sure to secure the IP removal unit in place. If not, the up-down IP removal unit may
fall down, because the worm gear becomes free when MB3 is removed.
(2)
Cable tie
(3) Remove the cable clamp.
(4) Disconnect the connector (CNMB3).
(5) From the rear of the machine, remove the two MB3 retaining screws, and take out
MB3.
(6) Remove the timing belt.
(7) Put a new timing belt onto the timing belt wheel.
(8) From the rear of the machine, secure MB3 in place with two screws.
(9) Put the cable clamp back where it was.
(10) Connect the connector (CNMB3).
(11) Cut the cable tie that retains the IP removal unit in place.
009-051-04
05.31.99 FM2460
3
4
5
6
Timing belt
wheel
Pan-head M4x10, x2
CR-IR341 Service Manual
7
8
9
10
(1)
Timing belt
IP removal unit
(6), (7)
Timing belt
(3), (4)
CNMB3
(5)
MB3
FR1H1195.EPS
MC - 77
FR1H1072.EPS
MC - 77