FujiFilm CR-IR341 Service Manual page 214

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
5.10
Empty Cassette Ejection
(1) Overall/General Sequence
Fig. 5.10.1 Empty Cassette Ejection
Preprocessing: Fig. 5.1.1
Same as
in the cassette
ejection routine
descried in
Fig. 2.11.1.
Table 5.10.1a List of I/O Names Used in the Flow and Drawing
Topmost
I/O name
shelf
in flow
position
SA1
SA1
SA2
SA2
LEDT1
LEDT1
LEDT2
LEDT2
009-051-04
05.31.99 FM2460
3
4
5
6
START
Is SA1 opened?
Y
N
Message output
"Unload cassette."
Cassette hold release
Fig.2.2.2
LEDT2 is turned on
Cassette removal
Is SA1 opened?
N
Y
DELAY TA13
Is SA2 opened?
N
Y
LEDT2 is turned off
LEDT1 is turned on
Empty cassettes in all shelves are ejected?
N
Y
END
I/O name
2nd shelf
3rd shelf
4th shelf
position
position
position
SA5
SA9
SA13
SA6
SA10
SA14
LEDT3
LEDT5
LEDT7
LEDT4
LEDT6
LEDT8
TR1H0088.EPS
CR-IR341 Service Manual
7
8
9
10
The cassette removable
condition is indicated.
Operation performed
by the operator
Instant return of
cassette is supported.
FR1H0438.EPS
Table 5.10.1 Operation Parameters
Symbol
Description
Wait for cassette OUT
TA13
MD - 151
Value
0.5 sec