(6) Slightly pull the cassette to ensure that the cassette cannot be removed.
(7) Select "2. DRIVE".
(8) Enter the number of the solenoid selected at step (5).
The actuator selected is driven.
Example: 2. DRIVE is selected
6:SVB1
11:FFM
INPUT THE NUMBER OF SENSOR
1 - 11 : 1
0.QUIT
1.NUMBER
2.DRIVE
3.STOP
MU:AUT > 2
1:SOLA1 2:SOLA2 3:SOLA3 4:SOLA4
6:SVB1
11:FFM
INPUT THE NUMBER OF SENSOR
1 - 11 : 1
0.QUIT
1.NUMBER
2.DRIVE
3.STOP
MU:AUT >
(9) Pull out the cassette.
(10) Repeat steps (2) through (9) several times to ensure that there is nothing abnormal.
If anything abnormal is found with a solenoid, replace it and then check it again.
(11) Select "0. QUIT".
The operation is exited to return to "2. ACTUATOR" menu.