(8) Rotate the MB3 timing belt counterclockwise, as seen from the rear, until the actuator
blocks the SA17 light path.
When the light path is blocked, a message indicative of the sensor CLOSE status
appears on the operation panel.
(9) Rotate the MB3 timing belt clockwise, as seen from the rear, until the actuator does
not block the SA17 light path.
When the light path is no longer blocked, a message indicative of the sensor OPEN
status appears on the operation panel.
CAUTION
Be sure to position the actuator below SA17. If the machine is booted up with the actuator
positioned above SA17, the home position (HP) cannot be detected during initialization, so that
the IP removal unit hits the top plate of the elevation unit, probably damaging the machine.
(10) Repeat steps (7) through (9) several times.
If anything abnormal is found, such as no message displayed, then replace the sensor.
(11) Repeatedly select "0. QUIT".
M-Utility is exited.
Replacement
(1) Disconnect the connector (S17).
(2) Remove the elevation HP sensor (SA17).
"3.5 Sensors Not Requiring Adjustment of Its Mounting Position" in Machine
Description
(3) Install a new sensor.
(4) Connect the connector (SA17).
(5) Check SA17.
" Checks"
009-051-04
05.31.99 FM2460
3
4
5
6
Bracket
Bracket
Latch
(2)
SA17
(1)
Connector