13A2 Cassette Hold Release Error
Cause
Because the IP is returned into the cassette, SOLA1-4 are turned ON to release the
cassette hold, but SA3, 7, 11, or 15 does not turn OFF, so that the cassette hold
cannot be detected.
Supplementary explanation
In order to release the cassette hold, SOLA1-4 are turned ON.
As SOLA1-4 turn ON, the hold pin is withdrawn, and SA3, 7, 11, and 15 turn OFF, so
that the hold pin release is detected.
This error occurs when either SA3, 7, 11, or 15 does not turn OFF (a shelf where the
cassette hold cannot be released) during the above operation.