(4) Make sure that the socket of the PMT08A board is fitted into the socket of the light-
collecting guide and that the sockets of the PMT08A board and light-collecting guide
are parallel to each other.
If anything abnormal is found, remove the single screw that retains the light-collecting
guide retaining bracket. While holding the light-collecting guide, push in the PMT08A
board assembly, and adjust the position of the sockets of the PMT08A board and light-
collecting guide as appropriate.
After the brace is fixed with the screw, verify again that the sockets are parallel, and
then proceed to the next step.
CAUTIONS
• Wear gloves when handling the light-collecting guide assembly. If it is handled with
bare hands, the light-collecting guide assembly may be soiled or damaged.
• Wear a mask to prevent saliva from sputtering around on the light-collecting guide.
Parallel
(5) Connect the connectors (CN1, 2 to the PMT08A board).
(5)
PMT08A
board
(6) Push in the scanner unit.
"7. Pulling Out and Pushing In Scanner Unit"
009-051-04
05.31.99 FM2460