Procedures for installing the timing belt
(1) Install the timing belt.
(2) Loosen the tensioner retaining screw, and flick the timing belt two or three times with
your fingers.
The tensioner is pulled by the spring force to the state where the tension of the
timing belt is adjusted.
(3) Secure the tensioner in place by tightening the screw.
FRONT
positioning conveyor
009-051-04
05.31.99 FM2460
3
4
5
6
Vertical conveyor
Before-side-
Erasure conveyor
CR-IR341 Service Manual
7
8
9
10
Tighten
screw
Timing belt
Before-side-
positioning conveyor
MD - 10
Vertical conveyor
Timing belt
Tensioner
Before-side-
positioning
conveyor
Tensioner
Tighten
screw
Erasure conveyor