C
1
2
ONTENTS
5.3
IP Search
(1) Overall/General Sequence
Fig. 5.3.1 IP Search (1/3)
SB2 CLOSE &
SC1 OPEN &
(SF1 | SF2 | SF3 CLOSE)
SB2 CLOSE &
SC1 & SF1 & SF2 & SF3
OPEN
Movement
of feed/load
conveyance
remaining IP
Fig. 5.3.3
SB2 OPEN &
SB2 OPEN &
(SC1 | SF2 CLOSE)
SF3 CLOSE
Positioning of
feed/load
conveyance
remaining IP
Fig. 5.3.6
IP found
(after search)
(SF1 | SF2 CLOSE)
or
SF1 & SF2 & SF3 CLOSE
009-051-04
05.31.99 FM2460
3
4
5
6
Preprocessing: Fig. 5.1.1
Search for
cassette inlet IP
Fig. 5.3.2
IP found
No IP
(after search)
SB2 is CLOSE
SB2 & SC1
OPEN &
(SF1 | SF2 | SF3 CLOSE)
SB2 CLOSE &
SC1 OPEN &
(SF1 | SF2 | SF3 CLOSE)
Movement
of feed
conveyance
remaining IP
Fig. 5.3.4
SB2 CLOSE
IP found
(after search)
(SF1 | SF2 | SF3 CLOSE)
(SF1 | SF2 | SF3 CLOSE)
Fig. 5.3.7
IP found
(after search)
To Fig. 5.3.1 (3/3)
A
Processing of before-reading conveyance remaining IP
CR-IR341 Service Manual
7
8
9
10
START
Is SB2 closed?
N
Y
(NOTE) Three steps marked by * in the figure,
"search for feed conveyance remaining
IP," "processing of before-reading
conveyance remaining IP," and
"search for load conveyance remaining
IP," are activated concurrently.
SB2 is OPEN
SB2 OPEN &
SC1 CLOSE
Search for feed
conveyance
remaining IP
Fig. 5.3.5
SB2 OPEN
No IP
SB2 & SC1 &
SF1 & SF2 & SF3
OPEN (NOTE)
Fig. 5.3.1
(2/3)
* Search for
feed conveyance
remaining IP
Fig. 5.3.5
SB2
SB2
OPEN
CLOSE
IP found
(after
search)
No IP
To Fig. 5.3.1 (2/3)
Processing of
after-reading
conveyance
B
remaining IP
MD - 120
Processing of
cassette inlet
IP
Processing of
feed/load
conveyance IP