11.4 4. SENSOR: Operation Tests on Sensors
It is used during checking of the sensors. Operation tests on the sensors are performed.
When the sensor status changes, an audible beep alert is generated. If anything abnormal is
detected, an error message is displayed.
1. NUMBER: To Display Sensor No. List
Execute "1. NUMBER".
A list of sensor numbers is displayed as follows.
1:SA1
6:SA6
11:SA11 12:SA12 13:SA13 14:SA14 15:SA15
16:SA16 17:SA17 18:SB1
21:SB4
26:SD2
31:SF3
2. MONITOR: To Monitor Individual Sensor
(1) Select "2. MONITOR".
The following message appears.
1:SA1
6:SA6
11:SA11 12:SA12 13:SA13 14:SA14 15:SA15
16:SA16 17:SA17 18:SB1
21:SB4
26:SD2
31:SF3
INPUT THE NUMBER OF SENSOR.
1 - 34 :
Meaning: Enter the number of sensor.
(2) Type in the number of sensor.
Sensor monitoring is initiated, and, when the sensor status changes, the following
message appears.
XXXXX ----> Open
XXXXX ----> Close
Meaning: Sensor XXXXX becomes OPEN.
009-051-05
11.20.99 FM2638
13
14
15
16
2:SA2
3:SA3
7:SA7
8:SA8
22:SB5
23:SC1
27:SD3
28:SE1
32:SF4
33:SZ2
2:SA2
3:SA3
7:SA7
8:SA8
22:SB5
23:SC1
27:SD3
28:SE1
32:SF4
33:SZ2
Sensor XXXXX becomes CLOSE.
CR-IR341 Service Manual
17
4:SA4
5:SA5
9:SA9
10:SA10
19:SB2
20:SB3
24:SC2
25:SD1
29:SF1
30:SF2
34:SZ3
4:SA4
5:SA5
9:SA9
10:SA10
19:SB2
20:SB3
24:SC2
25:SD1
29:SF1
30:SF2
34:SZ3
MU - 97