FujiFilm CR-IR341 Service Manual page 641

Fuji computed radiography
Table of Contents

Advertisement

1
2
C
ONTENTS
Checks after LD Assembly Replacement
CAUTION
Before removing the covers, be sure to turn OFF the high-voltage switch. If the machine is
powered ON with any of the covers removed, the photomultiplier will be damaged.
(1) Power ON the machine.
Make sure that the machine starts up normally.
If the machine does not start up normally, take the following steps to perform self-
diagnostics of Scanner Utility.
1. Touch the U-Utility button.
The screen switches to the U-Utility mode.
2. Touch the upper left and upper right corners of the operation panel in sequence.
M-Utility is activated.
3. Select "5. SCANNER UTILITY", "3. LASER", and "2. ON" in sequence.
Verify that no error occurs.
(2) Check image performance.
"3.2 Checking Image" in Preventive Maintenance Volume
009-051-04
05.31.99 FM2460
3
4
5
6
CR-IR341 Service Manual
7
8
9
10
MC - 150
MC - 150