11.2 Checking and Replacing Grip Release HP Sensor (SE1)
If anything abnormal is found during checks, replace SE1 and then check it again.
Checks
CAUTION
Before removing the covers, be sure to turn OFF the high-voltage switch. If the machine is
powered ON with any of the covers removed, the photomultiplier will be damaged.
(1) Power ON the machine.
(2) Touch the U-Utility button.
The screen switches to the U-Utility mode.
(3) Touch the upper left and upper right corners of the operation panel in sequence.
M-Utility is activated.
(4) Select "6. MECHANICAL UTILITY" and "3. SENSOR" in sequence.
(5) Select "2. MONITOR".
A list of sensor Nos. is displayed.
(6) Type in the number of SE1.
The machine is now in the sensor monitoring enabled state.
(7) Using a rule or the like, block the SE1 light path.
A message indicating that the sensor is CLOSE is displayed on the operation panel.
(8) Remove the rule that blocks the SE1 light path.
A message indicating that the sensor is OPEN is displayed on the operation panel.
(9) Repeat steps (7) and (8) several times.
If no message appears, it is likely that the sensor is faulty. So, replace the sensor.
After-reading
conveyor
(10) Repeatedly select "0. QUIT" to exit M-Utility.
A message prompting you to reset the machine is displayed.