FujiFilm CR-IR341 Service Manual page 422

Fuji computed radiography
Table of Contents

Advertisement

1
2
C
ONTENTS
13AE
Preliminary steps:
1. Power OFF the machine and remove the covers.
2. Turn OFF the high-voltage switch (S1) of the SCN08A board.
3. Check for any foreign matter (lead character marking chips, etc.) inside the machine, and if any, remove it all.
Power ON
Does the exposure information of
the patient correspond to the
image processing conditions of
the machine?
Power OFF
Is the IP normal?
Make sure that the barcode
indication is free of soil.
Power OFF
Is the BCR reading face normal?
Make sure that the BCR reading
face is free of soil.
Restart the machine
and check its status.
009-051-04
05.31.99 FM2460
3
4
5
6
N
Y
N
Y
N
Y
CR-IR341 Service Manual
Check the exposure information of the patient and the image
processing of the machine, and execute it again.
Clean the IP barcode indication.
Clean the BCR reading face.
MT - 153