10.10 Replacing Driven Shaft Grip Motor (MZ3)
Removal
(1) Pull out the scanner unit.
"7. Pulling Out and Pushing In Scanner Unit"
(2) Remove the side-positioning conveyor.
"8.1 Removing and Reinstalling Side-Positioning Conveyor"
(3) Remove the FAST-ON terminals (MZ2(+), MZ2(-)).
(4) Remove the right-hand side transparent cover.
(5) Remove the MZ3-attached bracket as a whole.
(6) Remove the ring and cam from the MZ3 shaft.
(7) Remove MZ3.
Subscanning
Positioning
bracket
(8) Reverse the removal steps to reinstall MZ3.
When installing the MZ3 bracket, screw it down while pressing it against the position-
ing bracket.
009-051-04
05.31.99 FM2460
3
4
5
6
(4)
Right-hand side transparent cover
BR M4x8, x2
unit
(6)
Ring
Hex socket head
setscrew
M3x6
CR-IR341 Service Manual
7
8
9
10
Left side of subscanning unit
Rear side plate
(5)
Bracket
BR M4x8, x4
(6)
Cam
SZ3
MC - 191
(7)
MZ3
Sems M3x6 x3
(3)
FAST-ON terminal (-)
(3)
FAST-ON terminal (+)
FR1H1021.EPS
MC - 191