4.4
Checking and Replacing Cassette Ejection Sensors (SA1,
5, 9, 13)
Checks
(1) Place the machine in a sensor monitoring state so that SA1, 5, 9, and 13 can be
monitored.
"2.2 Checking Various Sensors"
(2) Set a cassette into the shelf in place.
A message indicating that the sensor is CLOSE appears on the operation panel.
Cassette
(3) Pull out the cassette.
A message indicating that the sensor is OPEN appears on the operation panel.
(4) Repeat steps (2) and (3) several times to ensure that the messages appear as appro-
priate.
If anything abnormal is found, such as no message displayed, then replace the sensor
checked, and then check it again.
(5) Select "0. QUIT".
The monitoring is exited to return to "3. SENSOR" menu.
009-051-04
05.31.99 FM2460
3
4
5
6
CR-IR341 Service Manual
7
8
9
10
Cassette set unit
MC - 32
SA1
SA5
SA9
SA13
FR1H4045.EPS