FujiFilm CR-IR341 Service Manual page 452

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
Blank image
No. 107
• A white blank image is generated.
Uniformly exposed image
No. 108
with midtone
• A uniformly exposed image with midtone density
(gray) is generated.
009-051-04
05.31.99 FM2460
3
4
5
6
CR-IR341 Service Manual
<Probable Causes>
• The high-voltage switch (S1) of the SCN08A
board and the software switch are in the OFF
position.
• The PMT08A board is faulty.
<Check and Reference>
• Set the SCN08A board's high-voltage switch (S1)
and the software switch in the ON position.
"Checking SCN08A Board"
"Checking PMT08A Board"
<Probable Causes>
• The connector (CN2) is not connected to the
PMT08A board.
• The connector (CN5) is not connected to the
SCN08A board.
• Quasi-reading mode is set.
<Check and Reference>
• Connect the unplugged connector to the
PMT08A board or SCN08A board.
• Cancel the quasi-reading mode.
MT - 183
FR1H1856.EPS
FR1H1857.EPS
MT - 183