FujiFilm CR-IR341 Service Manual page 273

Fuji computed radiography
Table of Contents

Advertisement

1
2
3
4
C
ONTENTS
Code
Name
0813
DSP11 memory read error
0814
DSP10 checksum error
0815
DSP10 self-diagnostics error
0816
DSP10 image processing error
0817
DSP11 device initialization error
0818
DSP11 µ load error
0819
DSP11 memory write error
081A
DSP11 memory read error
081B
DSP11 checksum error
081C
DSP11 self-diagnostics error
081D
DSP11 image processing error
081E
DSP20 device initialization error
DSP20 µ load error
081F
0820
DSP20 memory write error
009-051-04
05.31.99 FM2460
5
6
Significance/Occurrence Condition
Read from the DSP device cannot be done.
DSP µ checksum error
DSP µ self-diagnostics error
DSP µ image parameter error
The DSP device cannot be opened.
The DSP µ is not yet registered in the HD.
Write to the DSP device cannot be done.
Read from the DSP device cannot be done.
DSP µ checksum error
DSP µ self-diagnostics error
DSP µ image parameter error
The DSP device cannot be opened.
The DSP µ is not yet registered in the HD
Write to the DSP device cannot be done.
CR-IR341 Service Manual
Detail: Denotes the detail information format.
Probable Cause/Remedy
Check the IMG08A board.
• Check the IMG08A board.
• Register the DSP µ in the HD again.
• Check the IMG08A board.
• Register the DSP µ in the HD again.
Check the image parameter.
Check the IMG08A board.
Register the DSP µ in the HD again.
Check the IMG08A board.
Check the IMG08A board.
• Check the IMG08A board.
• Register the DSP µ in the HD again.
• Check the IMG08A board.
• Register the DSP µ in the HD again.
Check the image parameter.
Check the IMG08B board.
Register the DSP µ in the HD again.
Check the IMG08B board.
MT - 14
Detail
Refer to
I-7
I-7
I-7
I-7
I-7
I-7
I-7
I-7
I-7
I-7
I-7
I-7
I-7
I-7
TR1H1228 .EPS
MT - 14