(5) Remove the rubber rollers.
Dislodge the shaft on one side of the four rubber rollers (right as seen from the rear)
so that the vertical conveyor guide can be removed.
(6) Remove the elevation changeover guide.
"5.14 Replacing and Adjusting Elevation Changeover Guide"
(7) Remove the vertical conveyor guide.
(8) For reinstallation, reverse the removal steps.
To install the vertical conveyor guide, tighten screws while pushing down the guide.
009-051-04
05.31.99 FM2460
3
4
5
6
CR-IR341 Service Manual
7
8
9
10
(7)
Vertical conveyor guide
BR M4x8, x2
MC - 85
(6)
Elevation
changeover guide
(5)
Rubber roller x4
FR1H1076.EPS