FujiFilm CR-IR341 Service Manual page 709

Fuji computed radiography
Table of Contents

Advertisement

C
11
12
ONTENTS
12. Erasure Conveyor
This section describes procedures for removing and reinstalling the erasure lamp and
erasure conveyor, and procedures for checking and replacing parts of the erasure conveyor
to be replaced.
Unless otherwise specified, it is assumed that covers that should be removed prior to checks
and replacement have been removed.
"3. Removing and Reinstalling Covers"
12.1 Removing and Reinstalling Erasure Lamp Assembly
Removal
(1) Disconnect the connectors (ACOUT1, ACOUT2, DCOUT4, DCOUT5).
(2) Remove the single screw that retains the erasure lamp assembly.
(3) Pull out the erasure lamp assembly horizontally.
CAUTION
Carefully handle the erasure lamp. If it is hit, the erasure lamp may be broken, so you
may get injured.
Erasure conveyor
009-051-04
05.31.99 FM2460
13
14
15
16
CR-IR341 Service Manual
(1)
Power supply
DCOUT4
ACOUT1
(3)
MC - 218
DCOUT5
ACOUT2
(2)
Erasure lamp assembly
BR M4x8
FR1H1180.EPS
MC - 218