FujiFilm CR-IR341 Service Manual page 427

Fuji computed radiography
Table of Contents

Advertisement

1
2
C
ONTENTS
13E2 Feed IP Dropped/13E5 Load IP Dropped
Cause
13E2: While the IP is being sucked and conveyed by the suction cups, SB5 is turned OFF.
13E5: While the IP that has been loaded is being sucked and conveyed into the cassette,
SB5 is turned OFF.
Supplementary explanation
13E2:
The IP in the cassette is sucked and conveyed so that the IP is fed to the grip roller.
While the IP is sucked by the suction cups, SB5 is ON.
This error occurs when SB5 is turned OFF while the IP is being moved by predeter-
mined pulses to the grip rollers during the above operation.
13E5:
The IP that has been loaded is sucked and conveyed back into the cassette.
While the IP is sucked by the suction cups, SB5 is ON.
This error occurs when SB5 is turned OFF while the IP is being moved by predeter-
mined pulses to the grip roller during the above operation.
Sensor related
Error-related I/O locations
Error-related system block diagram
AC IN
DC OUT2
JPS-1
009-051-04
05.31.99 FM2460
3
4
5
6
ON: Close, OFF: Open
SNS08A
CN
CN13
CN4
DRV03D
CR-IR341 Service Manual
SB5
Suction sensor
FR1H0128.EPS
CNB1
CNB2
FR1H0129.EPS
MT - 158
MT - 158