FujiFilm CR-IR341 Service Manual page 873

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
10.7 7. SHADING/SENSITIVITY: Setting Shading and Sensitivity
Correction Data
The shading and sensitivity correction data is recorded, calculated, and set. Note, however,
that because the setup value in the HD is not overwritten, the data so obtained is lost upon
resetting.
Writing the settings to the hard disk is done in "8. DATA MANAGEMENT."
For more detail about the procedures for shading and sensitivity correction, see the Checks,
Replacement and Adjustment Volume.
"9.2 Shading/Sensitivity Correction" in Checks, Replacement and Adjustment Volume
1. REC MODE: To Make Correction Data Recordable
The shading, polygon, and sensitivity correction data is made recordable.
Execute "1. REC MODE".
The correction data is now recordable, with the following message displayed.
THE SPECIAL MODE TO RECORD ADJUSTMENT DATA IS SET.
Meaning: The correction data recording mode is set.
NOTE
REC MODE (recording of correction data) is effective for a single reading. To record
correction data repeatedly, execute "1. REC MODE" again.
To Test to Check If Image Size is as Specified
: 2.CALCULATION : 1.SHADING, POLYGON AND SENSITIVITY
: 2.CALCULATION : 2.SHADING, POLYGON AND SENSITIVITY FOR BAD CONDITION
: 2.CALCULATION : 6.SENSITIVITY ONLY
(1) Select "1. SHADING, POLYGON AND SENSITIVITY".
When the recorded data is correct, a prompt for input of dosage value appears as
follows.
INPUT DOSAGE VALUE.
0.50 - 9.99 :
The unit is in mR.
If the recorded data is incorrect, the following message appears.
THE DATA IS RECORDED WITH INCORRECT SIZE.
Meaning: The data has been recorded with wrong size.
(2) Enter the dosage value.
The calculation of the correction data is initiated, with the following message dis-
played.
CALCULATION IS IN PROGRESS.
RESULT - OK (or RESULT - ERROR)
Meaning: Correction data calculation is now being performed.
009-051-05
11.20.99 FM2638
3
4
5
6
CR-IR341 Service Manual
7
8
9
10
MU - 82